Record and assess one measured Silt Density Index (SDI) replicate against the selected method. No RO-feed acceptance decision is made.

Seawater desalinationStage 02 Stage 05
Status
Available
RO PRETREATMENT · LABORATORY RECORD

SDI Test Assessment

Enter one timed 500 mL test record and review its method qualification.

Method not selected
SDI TEST MEASUREMENTS

Three recorded times

T · ti · tf

Use the measured interval and collection times. The calculator records one replicate only.

TEST CONDITIONS

Method record

App record cites Rev. 20; the linked manual is Rev. 19. Rev. 20 is not independently confirmed. Manual reference

SAMPLE RECORD

Identify the replicate

Recorded separately; it does not calculate SDI.
Additional test details

Calculate a complete test record to enable its report.

01

Engineering Basis

Methodology

Calculates SDI_T = (1 - ti/tf) * 100/T from one recorded 500 mL timing pair when inputs match the calculator's declared Rev. 20 (August 2026), Section 2.5.1 record. The current English manual linked from DuPont's product page identifies as Rev. 19 (February 2026). Targeted official-source checks did not independently confirm Rev. 20; this does not establish that no such revision exists. A matching result confirms agreement with the app record only. Plugging is reported separately.

Scope / Limitations

  • Calculates method-qualified SDI only when the recorded timing, 47 mm holder, 0.45 µm filter, 500 mL sample, pressure, interval and completion conditions match DuPont FilmTec Manual Rev. 20, Section 2.5.1. Turbidity is separate; replicates are not combined. This is not RO-feed acceptance or membrane-safety advice.